Sterilization and test cycles table
B
delicate
B
standard
B
prion
S
ash
Cycle type
EN13060
B
B
B
S
Helix & B.D.
Test
Air
LeakageTest
Sterilization
Temperature
121°C 134°C 134°C 134°C
134°C
-
Sterilization
Pression
1.1 Bar
2.15 Bar
2.15 Bar
2.15 Bar
2.15 Bar
-
Sterilization
phase
18 Minutes 4 Minutes 20 Minutes 4 Minutes
3.5 Minutes
-
Drying phase
0-15-30 -
ECO
Minutes
0-15-30-
ECO
Minutes
0-15-30-
ECO
Minutes
0-5-10-
ECO
Minutes
-
-
Total cyle
time* SST1700B
35 Minutes
22 Minutes
38 Minutes
16 Minutes
17 Minutes
-
Total cyle
time* SST2200B
37 Minutes
25 Minutes
41 Minutes
17 Minutes
18 Minutes
-
*warm status+small load +
ECO
drying
Unpacked, packed or double packed
Unpacked
Emtpy chamber
or test PCD’s/indicators
Solid
Yes
Yes
Yes
Yes
Small porous
Yes
Yes
Yes
No
Porous
Yes
Yes
Yes
No
Hollow type A
Yes
Yes
Yes
No
Hollow type B
Yes
Yes
Yes
Yes
LIMITS OF LOAD (Gross weight)
Maximum solid
load
6 Kg (23lts)
4.5 Kg (17lts)
6 Kg (23lts)
4.5 Kg (17lts)
6 Kg (23lts)
4.5 Kg (17lts)
6 Kg (23lts)
4.5 Kg (17lts)
Not suitable for sterilization
Maximum
porous load
2 Kg (23lts)
1.5 Kg (17lts)
2 Kg (23lts)
1.5 Kg (17lts)
2 Kg (23lts)
1.5 Kg (17lts)
-
SST
Sterilizers
are equipped with a
powerful vacuum
pump
that assures the performances of the air removal sy-
stem through
fractionated vacuum
.
SST
Sterilizers
are provided with an
intelligent drying
mode
called
ECO
Drying
that
automatically
evaluates the load volume
and decides the correct drying
time. The
ECO
Drying reduces
drammatically
the total cycle
time and reduces the electricity consumption as well.
The
P.I.D. control system
assures the
maximum accuracy
and
“Automatic Evaluation System”
controls the achieving
of the
EN13060 requirements
in every cycles.
CYCLE
IN LESS THAN
25 MINUTES
1,2,3,4 6,7,8