CBM - Catalog SST 2200-1700 EN

5 Sterilization and test cycles table B delicate B standard B prion S ash Cycle type EN13060 B B B S Helix & B.D. Test Air LeakageTest Sterilization Temperature 121°C 134°C 134°C 134°C 134°C - Sterilization Pression 1.1 Bar 2.15 Bar 2.15 Bar 2.15 Bar 2.15 Bar - Sterilization phase 18 Minutes 4 Minutes 20 Minutes 4 Minutes 3.5 Minutes - Drying phase 0-15-30 - ECO Minutes 0-15-30- ECO Minutes 0-15-30- ECO Minutes 0-5-10- ECO Minutes - - Total cyle time* SST1700B 35 Minutes 22 Minutes 38 Minutes 16 Minutes 17 Minutes - Total cyle time* SST2200B 37 Minutes 25 Minutes 41 Minutes 17 Minutes 18 Minutes - *warm status+small load + ECO drying Unpacked, packed or double packed Unpacked Emtpy chamber or test PCD’s/indicators Solid Yes Yes Yes Yes Small porous Yes Yes Yes No Porous Yes Yes Yes No Hollow type A Yes Yes Yes No Hollow type B Yes Yes Yes Yes LIMITS OF LOAD (Gross weight) Maximum solid load 6 Kg (23lts) 4.5 Kg (17lts) 6 Kg (23lts) 4.5 Kg (17lts) 6 Kg (23lts) 4.5 Kg (17lts) 6 Kg (23lts) 4.5 Kg (17lts) Not suitable for sterilization Maximum porous load 2 Kg (23lts) 1.5 Kg (17lts) 2 Kg (23lts) 1.5 Kg (17lts) 2 Kg (23lts) 1.5 Kg (17lts) - SST Sterilizers are equipped with a powerful vacuum pump that assures the performances of the air removal sy- stem through fractionated vacuum . SST Sterilizers are provided with an intelligent drying mode called ECO Drying that automatically evaluates the load volume and decides the correct drying time. The ECO Drying reduces drammatically the total cycle time and reduces the electricity consumption as well. The P.I.D. control system assures the maximum accuracy and “Automatic Evaluation System” controls the achieving of the EN13060 requirements in every cycles.